The automatic micro moisture analyzer meets the requirements of GB/T7600, GB/T6283, SH/T0246, SH/T0255 and other methods and standards. Karl Fischer (Coulomb) method was used to determine the content of trace water in liquids with different properties. It has the advantages of high measurement accuracy, fast speed, stable and reliable measurement data, and is widely used in petroleum, chemical, electric power, environmental protection, medicine and other departments. It can quickly determine the water content of chemical, petroleum, pharmaceutical, pesticide and other products such as alcohols, oils, lipids, ethers, esters, acids, alkanes, benzene, amines, organic solvents, pesticides, phenols, pharmaceutical raw materials, etc.
Technical features:
The 64 bit embedded microprocessor is used as the main control core and embedded in the mini operating system
Constant current detection, high precision, fast measurement speed, background compensation function, stable and reliable
Double circuit balanced titration, with smaller test result error and higher sensitivity
Electrolytic cell adopts new waterproof sealing design, with short balancing time and longer electrolyte service time
The host is equipped with automatic discharge and filling functions, greatly improving the working environment
Real time electrolysis curve graph to observe reagent status at any time
Large screen touch LCD display, simple operation and friendly interface
Full number keyboard, supporting 6 calculation formulas, convenient and fast data calculation
The whole machine adopts engineering plastic shell, which is light and beautiful
Built in large capacity storage, more convenient for recording and querying measurement data
Technical parameters:
Electrolytic current |
0 ~ 400mA |
Measuring range |
3 μ G~100mg (water) |
Resolution |
0.1 μ g |
Precision |
3 μ G water ± 20%, 10 μ G water ± 10%, 100 μ G water ± 5%, 100 μ ± 0.5% above g water (excluding injection error) |
Display mode |
Chinese LCD |
Printer thermal type |
36 characters, Chinese character output |
Ambient temperature |
5 ℃~ 40 ℃ |
Power supply voltage |
AC 220V ± 10% 50Hz ± 10% |
Power consumption |
<40W |
Overall dimension |
300mm × 290mm × 200mm |
Weight |
3.7kg |