Pensky-Martens Flash Point Tester SL-OA10

Pensky-Martens Flash Point Tester SL-OA10 Description

Product description:

This instrument is designed and manufactured according to the requirements stipulated in the national standard GB/T261-2021;ASTM D93 "Determination of Flash Point Pensky-Martens Closed Cup Method". It is suitable for determining the flash point of petroleum products with a closed cup, under the specified conditions, heated to the lowest temperature of its vapour and air mixture in contact with the flame when the flash fire occurs, i.e. the closed cup method of flash point, in accordance with the method specified in the standard.


Technical features:

1. SL-OA10 automatic closed cup flash point tester is used to determine the closed flash point of petroleum products.

2.8-inch pure industrial true colour touch screen, Chinese and English interface to switch at will, WindowsCE embedded operating system.

3.With self-diagnostic function to check whether there is a fault and display its status.

4.Automatically correct the effect of atmospheric pressure on the test and calculate the correction value.

5.Standard RS485 interface, can also be connected to the USB2.0 interface, the instrument parameters can be set and upload data to the computer.The lower computer stores 120 sets of historical data.

6.Ignition mode manual and automatic optional.

7.detection, open the lid, ignition, printing, fully automatic completion.

8.using the original imported Renesas microcontroller. Germany imported PT1000 high-precision temperature sensor.   

9.with the prevention of fire over-temperature automatic stop working function.

10.built-in high-precision flash point detection sensor, detection time up to milliseconds.

11.analogue tracking display temperature rise and test time function curve.


Technical parameters:

①Working power supply: AC220V±10%, 50Hz;
②Temperature measurement: range: room temperature ~ 350 ℃;
Repeatability: ≤2℃; reproducibility: ≤4℃;
Resolution: 0.1℃; precision: 0.5%;
③Base parameters: Rising speed: in line with GB/T261-2008 standards;
Ignition mode: electronic ignition, gas flame;
④ambient temperature: (10 ~ 40) ℃; relative humidity: ≤ 80%;
⑤Power consumption of the whole machine: not more than 500W.